Due to the internal non-engaging circular geometry of the Lifecore Stage-1 implant, there are no implant level analogs or impression copings. Restore the implant by torquing in an abutment and taking an abutment impression.
Titan Implants has solid COC abutments that can be torqued into the implant, along with matching solid abutment snap on transfer caps, analogs, and waxing copings.
The Titan solid COC abutment comes in two different crown retention shapes. One has a flat side (original Lifecore Stage-1 abutments had a flat side) with grooves and the t
he other is called the Titan Design which has a hex instead of a flat, offering better crown retention. Both work well.
Used in both anterior and posterior areas of the mouth for cement-retained crown and bridge restorations
The Titan solid abutment has ring grooves designed for mechanical restoration.
HOW TO CHOOSE A PREP HEIGHT:
- Three Prep Heights (4 mm, 5.5 mm, 7 mm) These graduated preparation heights give you precise control over the occlusal clearance and crown height.
- 4 mm: Ideal for limited interocclusal space or shallow preparations.
- 5.5 mm: Most commonly used for standard cases.
- 7 mm: Provides extra height for deep soft tissue or when more retention is needed. This range reduces the need for excessive chairside grinding or adding material, saving time and preserving abutment strength.
*Requires solid abutment driver part#ITI-SAD, Torque wrench 30Ncm
To fit original Straumann torque wrench:
SKU# ITI-SAD Solid Abutment Socket Wrench can be used with Straumann Torque Wrench Adaptor
SKU# IS-TWSHA:
OR
To fit original Nobel Biocare torque wrench:
SKU# ITI-SAD Solid Abutment Socket Wrench can be used with Nobel Biocare Torque Wrench Adaptor
SKU: NB-TWSHA:
See instruction videos for this system